Thermo Fisher Scientific's EXENT System, a mass spectrometry blood test, has been cleared by the US Food and Drug Administration (FDA) to help monitor multiple myeloma and Waldenstrom's macroglobulinaemia, the company announced on 24 November 2026. Until now, its US clearance covered only help with diagnosis. The FDA database lists a matching decision dated 20 November.
The system pairs the EXENT Analyser with the Immunoglobulin Isotypes (GAM) assay, both from The Binding Site, the Birmingham, UK, diagnostics business owned by Thermo Fisher. Ireland is not among the countries the company has named for the system.
- Cleared use: to aid monitoring of disease status and treatment response in multiple myeloma and Waldenstrom's macroglobulinaemia (WM).
- FDA record: the FDA database lists a second 510(k) for the assay, K261143, received 7 April 2026 and decided 20 November 2026. The company release gives no number.
- First clearance: K250159, as an aid to diagnosis, decided 17 October 2025.
- Europe: launched in 2023 with a CE mark under the EU's IVDR and monitoring in its intended use, the company says.
- Ireland: not among the 13 countries Thermo Fisher named in November 2025.
What has the FDA cleared?
Thermo Fisher says the FDA has cleared an expanded intended use for the assay: helping US clinical laboratories monitor disease status and treatment response in people with multiple myeloma or WM. It calls this a "first-of-its-kind FDA clearance".
The release gives no clearance number. The FDA's public database lists a second 510(k) for the assay, K261143, filed by The Binding Site Group Ltd of Birmingham and found "substantially equivalent" on 20 November 2026. The decision summary is not yet posted, so the exact wording of the new use is not public.
Stephen Harding, vice president and general manager, protein diagnostics, at Thermo Fisher, said: "This expanded clearance brings the precision of mass spectrometry beyond diagnosis and into ongoing monitoring, giving clinicians greater insight into disease status and treatment response."
What is an M-protein and why does monitoring matter?
Multiple myeloma is a cancer of plasma cells, the white blood cells that make antibodies. Myeloma UK explains that in myeloma these cells "produce a large amount of a single type of antibody (known as paraprotein or M-protein) which has no useful function". WM is a rare, slow-growing lymphoma whose abnormal cells make an antibody called IgM, according to Lymphoma Action.
Myeloma UK says a haematologist looks for the paraprotein level to fall or hold steady during treatment. Thermo Fisher says patients are typically monitored for life. Anyone with questions about their own results should talk to their GP or specialist.
How does the EXENT System work?
Mass spectrometry separates molecules by mass and charge. The FDA's decision summary for the first clearance describes two linked instruments: one prepares the serum sample with antibody-coated magnetic beads, and a MALDI-TOF mass spectrometer reads it. Results are combined with Binding Site's Optilite immunoglobulin tests to estimate the amount of M-protein.
Some myeloma medicines are themselves antibodies, and Thermo Fisher says they can interfere with conventional tests. The FDA summary says the system was trained to detect and flag three: daratumumab, isatuximab and elotuzumab.
An International Myeloma Working Group committee concluded in Blood Cancer Journal in 2021 that this type of mass spectrometry can replace an older electrophoresis method for diagnosing and monitoring myeloma. Its report covered the technique, not this product.
What evidence is public so far?
Thermo Fisher's release gives no performance figures for monitoring, a gap Targeted Oncology also noted. In November 2025 the company stated that the test "has not been evaluated for use in post-diagnostic patient monitoring of monoclonal gammopathies".
The published data come from the diagnostic clearance. The FDA summary reports that the test picked up 170 of 180 multiple myeloma cases (94.4%) and 63 of 64 WM cases (98.4%). It gave a correct negative result in 88.4% of 655 patients with other illnesses (the company's own summary says 87.8%). These figures describe diagnosis, not monitoring.
| Date | Event |
|---|---|
| 24 August 2023 | EXENT launched in Europe with IVDR certification |
| 17 October 2025 | K250159 cleared as an aid to diagnosis |
| 7 April 2026 | FDA receives second 510(k), K261143 |
| 20 November 2026 | K261143 found substantially equivalent |
| 24 November 2026 | Thermo Fisher announces monitoring clearance |
What it means for Ireland
The release does not mention Ireland. It says the system has tested thousands of patients in 13 countries over three years, without naming them. Thermo Fisher's November 2025 release listed the US, Australia, Belgium, Brazil, Canada, France, Germany, Italy, the Netherlands, New Zealand, Spain, Switzerland and the UK. Ireland is not on that list.
In the EU, Thermo Fisher says the system carries a CE mark under the In Vitro Diagnostic Regulation (IVDR). Its August 2023 launch release said the system was intended to aid diagnosis and the monitoring of multiple myeloma and WM, so monitoring was already part of the European product.
Thermo Fisher says it has more than 1,100 colleagues in Ireland, including over 500 in Cork making active pharmaceutical ingredients and about 400 at its PPD laboratory in Athlone. No source links these sites to EXENT or says the system is made in Ireland.
The National Cancer Registry Ireland counted an average of 392 new multiple myeloma cases a year in 2019 to 2021. In our view, Irish hospital laboratories now have a second regulator's decision to study, but no Irish adoption has been announced.
Frequently asked questions
Is the EXENT System available in Ireland?
Thermo Fisher has not named Ireland among the countries where the system is in clinical use. It says availability depends on each country's rules and tells laboratories to contact their local representative.
Does this change care for Irish patients?
Not directly. The clearance applies to the US. Patients should ask their GP or haematology team how their disease is monitored.
What does the EXENT test measure?
It measures monoclonal IgG, IgA and IgM, the M-proteins made by abnormal cells, in a serum sample.
Outlook
Watch for the FDA's decision summary for K261143, which should show the exact monitoring claim and the data behind it.
Sources
- Thermo Fisher Scientific, release on the expanded EXENT clearance, 24 November 2026
- FDA 510(k) database, K261143 and K250159, and K250159 decision summary, accessed 26 November 2026
- Thermo Fisher Scientific, US diagnostic clearance release, 12 November 2025, and EXENT launch with IVDR certification, 24 August 2023
- Targeted Oncology, FDA Expands EXENT Clearance to Myeloma and WM Monitoring, 25 November 2026
- Blood Cancer Journal, IMWG Mass Spectrometry Committee Report, 1 February 2021
- Myeloma UK, Paraproteins and free light chains, September 2020; Lymphoma Action, Waldenstrom's macroglobulinaemia, June 2025
- National Cancer Registry Ireland, Cancer Trends No 41, October 2024
- Thermo Fisher Scientific, Ireland, accessed 26 November 2026
Cover image: AI-generated illustration of a laboratory analyser carrying a Thermo Fisher Scientific label. It is not a photograph of the EXENT System, a Thermo Fisher laboratory or a real person.