Ireland
--:--IST
Latest
Astellas Expands Its 330 Million Euro Tralee Biopharma Facility with a Second Aseptic Filling Line to Double Drug-Product Capacity Xeolas Pharmaceuticals Opens 158,000 Sq Ft State-of-the-Art Baldoyle Facility to Scale Specialty Medicine Manufacturing Priya Life Science Partners with Fleming for the 9th Annual Corporate Compliance & Transparency in Life Sciences Summit in Zurich Ireland Has the Capital and the Lessons: Digital Project Management Is How They Become Delivery Dunbar Pharma Brings First Plant-Derived Dronabinol API to UK Market Through IPS Pharma Leveraging Priya Life Science as a Data Tracker: The Ultimate Use Case & Career Guide The €100K Reality Check: Why a Six-Figure Pharma Salary in Ireland Feels Different Than in Switzerland or Germany Ireland's €93.8 Billion Non-EU Pharma Export Engine: Trade Data, Destination Markets, and Economic Impact Astellas Expands Its 330 Million Euro Tralee Biopharma Facility with a Second Aseptic Filling Line to Double Drug-Product Capacity Xeolas Pharmaceuticals Opens 158,000 Sq Ft State-of-the-Art Baldoyle Facility to Scale Specialty Medicine Manufacturing Priya Life Science Partners with Fleming for the 9th Annual Corporate Compliance & Transparency in Life Sciences Summit in Zurich Ireland Has the Capital and the Lessons: Digital Project Management Is How They Become Delivery Dunbar Pharma Brings First Plant-Derived Dronabinol API to UK Market Through IPS Pharma Leveraging Priya Life Science as a Data Tracker: The Ultimate Use Case & Career Guide The €100K Reality Check: Why a Six-Figure Pharma Salary in Ireland Feels Different Than in Switzerland or Germany Ireland's €93.8 Billion Non-EU Pharma Export Engine: Trade Data, Destination Markets, and Economic Impact
Recruiting Not applicable

The Purpose of This Clinical Trial is to Evaluate the Accuracy of Blood Pressure Measurement of HiCardi M350 Compared to Blood Pressure Measured by Auscultatory Sphygmomanometer in Adult Volunteers (Including Those Meeting Blood Pressure Distribution Requirements) in Accordance With ISO 81060-2:2018

NCT07622498 · tracked via the Priya Life Science South Korea tracker
Phase
Not applicable
Started
2026-06-01
Last updated
2026-06-03

Condition(s) studied

No Restriction on Disease

Investigational drug(s) / intervention(s)

HiCardi M350

HiCardi M350: HiCardi M350 is a wearable patient monitoring device used to measure systolic and diastolic blood pressure. In this study, the device will be attached to each participant and calibrated according to the predefined procedure. Blood pressure values measured by HiCardi M350 will be compared with reference values obtained using an auscultatory sphygmomanometer and mechanical stethoscope by two independent observers.

Study summary

This prospective, single-center, single-arm pivotal clinical trial is designed to evaluate the accuracy of blood pressure measurements obtained using the HiCardi M350 patient monitor compared with measurements obtained using an auscultatory sphygmomanometer.

The study will enroll adult volunteers aged 19 years or older, including participants needed to satisfy the blood pressure distribution requirements of ISO 81060-2:2018. The planned enrollment is 100 participants, considering the ISO 81060-2:2018 minimum requirement of 85 participants and an estimated dropout rate of approximately 15%. The study is also designed to obtain at least 255 valid paired blood pressure values.

After written informed consent, participants will undergo screening assessments, including demographics, vital signs, medical and surgical history, prior and concomitant medications, pregnancy testing for women of childbearing potential, bilateral arm circumference measurement, eligibility assessment, and adverse event assessment. On the blood pressure test day, eligibility will be reconfirmed, an enrollment number will be assigned, HiCardi M350 calibration will be performed, and blood pressure measurements will be obtained using both HiCardi M350 and the reference auscultatory method.

The primary effectiveness endpoint is the difference between systolic and diastolic blood pressure values measured by HiCardi M350 and the reference sphygmomanometer under resting conditions. Accuracy will be evaluated according to ISO 81060-2:2018 using Criteria 1 and Criteria 2. Safety will be evaluated based on adverse events, adverse device effects, serious adverse events, serious adverse device effects, unanticipated adverse device effects, and device deficiencies.

Eligibility

Sex
ALL
Min age
19 Years
Max age
—
Healthy volunteers
Accepted
Inclusion Criteria: 1. Adults aged 19 years or older. 2. Individuals who visit Severance Hospital and voluntarily provide written informed consent after receiving and understanding sufficient explanation of the study. 3. Individuals with hypertension-range blood pressure may be included to meet the blood pressure distribution requirements. Participants receiving antihypertensive medication may be included only if the type and dose of medication have remained stable for at least 4 weeks before screening. Individuals with dose changes, medication additions, or medication discontinuation within 4 weeks before screening will be excluded. Exclusion Criteria: 1. Individuals in whom Korotkoff phase V cannot be clearly identified during auscultatory blood pressure measurement. 2. Individuals for whom blood pressure measurement is difficult or expected to be difficult. 3. Individuals with bilateral arm circumference greater than 42 cm. 4. Pregnant women. 5. Individuals currently participating in another clinical trial or who have participated in another clinical trial within 30 days before the screening date. 6. Individuals judged by the investigator to be inappropriate for participation in the clinical trial.

Primary outcome measure(s)

  • Difference between systolic and diastolic blood pressure measurements obtained using HiCardi M350 and a reference auscultatory sphygmomanometer — After completion of all participants' data collection (within approximately 7 months of enrollment period)
    The primary outcome is the difference between systolic and diastolic blood pressure values measured by HiCardi M350 and the reference auscultatory sphygmomanometer under resting conditions. Accuracy will be assessed separately for systolic and diastolic blood pressure according to ISO 81060-2:2018 Criteria 1 and Criteria 2.

Trial sites (1)

FacilityCityRegionStatus
Severance Hospital Seoul Seoul Recruiting
Official registry record

This page summarises publicly available registry data for informational purposes — not medical advice. Eligibility is determined by each study team; patients should discuss participation with their clinician.

View NCT07622498 on ClinicalTrials.gov ↗ ← All trials in South Korea